KR-8100PA/P |
SPECIFICATIONS |
|
KR-8100PA |
KR-8100P | |
| OBJECTIVE REFRACTOMETER MODE | ||
| Sphere |
-25 to +22D in 0.25D step (0.12D step available) |
|
| Cylinder |
0 to ±8D in 0.25D step (0.12D step available) |
|
| Axis |
1° to 180° in 1° step ( 5° step available) |
|
| Minimal pupil diameter |
2.0 mm Ø |
|
| Method of relaxation |
Automatic fogging |
|
| Fixation chart |
Picture chart |
|
| CORNEAL CURVATURE MODE | ||
|
Corneal curvature radius |
5.00mm ~ 10.00mm |
|
| Refraction index |
1.3375 |
|
| Corneal refraction |
67.5D ~ 33.75D |
|
| Corneal astigmatism |
0D ~ ± 10D |
|
| Corneal astigmatism axial angle |
1° ~ 180° |
|
| Measuring area |
3mm with 7.7 mm radius |
|
| MEASURING STEP | ||
|
Corneal curvature radius |
0.01 mm |
|
| Corneal refraction |
0.25D (0.12D step available) |
|
| Corneal astigmatism |
0.25D (0.12D step available) |
|
| Corneal axis angle |
1° (5° step available) |
|
| OTHERS | ||
| PD measurement range |
85 mm max. in 1 mm step |
|
| Measuring start |
Auto start / Manual start Selectable |
|
| Measuring alignment |
Auto/Manual Selectable |
Manual |
|
Corneal diameter measurement (Pupil diameter) |
Yes |
|
| Range ; 2 ~ 13 mm / step ; 0.25 mm | ||
| Measurement display |
TV monitor screen |
|
|
Measurement recording |
Built-in printer (Up to 10 measurements of each eye can be stored in memory) |
|
| Alignment |
Screen display |
|
| Vertex distance |
0, 12.0 and 13.75 mm (selectable) |
|
| Energy saving |
Automatic switch off when left unused after 10 minutes |
|
| IOL |
Special IOL switch to adjust to circumstances of IOL wearers |
|
| Power supply |
AC100-120, 220-240V |
|
| Output |
RS-232C, VIDEO-OUT |
|
| Weight |
21 kg. |
|
| Dimensions |